Labnet Z400 High Capacity Centrifuge User Manual
Page 13

4.4
Preselection of speed / RCF
You can preset the speed between 250 rpm amd 13500 rpm in steps of
250 rpm
4
1
5
3
2
When the centrifuge lid is open, you can preset the required speed or rcf with the
knob(2).
If the centrifuge lid is closed and during the run, the speed can be changed as
follows:
Press key "preset" (5), hold it and at the same time change speed with the knob
(2).
The preselected speed will be indicated in the speed display (4).
The centrifuge will accelerate only to the max. allowed speed of the inserted rotor.
The allowed max. speed for all rotors please see table under.
Z 400
Z 400
Z 400
Z 400
Z 400
Z 300
Z 300
Z 300
Z 300
Z 300
Rotor
max. Speed
Rotor
max. Speed
220.86 V01
3 500 rpm
220.72 V04
4 000 rpm
220.27 V02
4 000 rpm *
220.50 V05
3 500 rpm
220.50 V05
3 500 rpm
220.81 V05
3 500 rpm
220.81 V05
3 500 rpm
220.82 V01
6 000 rpm
220.82 V01
6 000 rpm
220.59 V06
13 500 rpm
220.59 V06
13 500 rpm
*with bucket 611.015 3500 rpm