TROUBLESHOOTING
1. Infusion Accuracy
To ensure infusion accuracy always use new syringe(s)
and measure syringe bore diameter and enter actual
dimensions in millimeters (mm) using the Custom Syringe
entry option. Additionally, make sure that the guide rods
and lead screw are properly lubricated.
KDS Legato 950 OEM Series Display Error
Messages
2. Motor Stalled
This indicates that the pusher block travel has been
impeded. This may be caused by the syringe plunger
hitting bottom, a kink in the tubing (occlusion), syringe
plunger binding or any situation requiring more force to
the head of the syringe than the pump is capable of
delivering. To restart the pump, remove the cause of
stalling and/or increase the force setting, then press
RUN. The pump will continue the interrupted procedure
where it stopped.
3. Out of Range
A value was entered or encountered in a pump program
that was beyond the pump’s limits.
4. Unit Won’t Power Up
Verify the the Power Source Switch on the rear of the
pump is set to DC-IN if powering through power supply
or RS-485 if powering in a chain.