Nominal Minimum/Maximum Flow Rates for Various Syringes.
(Actual Limits will vary depending on syringe manufacturer)
Syringe Size
Syringe ID
Minimum Rate
Units
Maximum Rate
Units
0.5 µl
0.103 mm
1.26000
pl/min
1.32611
µl/min
1 µl
0.1457 mm
2.52000
pl/min
2.65353
µl/min
2 µl
0.206 mm
5.10000
pl/min
5.30443
µl/min
5 µl
0.343 mm
14.1600
pl/min
14.7059
µl/min
10 µl
0.485 mm
28.2600
pl/min
29.4028
µl/min
25 µl
0.729 mm
63.9600
pl/min
66.4293
µl/min
50 µl
1.030 mm
127.680
pl/min
132.611
µl/min
100 µl
1.457 mm
255.480
pl/min
265.353
µl/min
250 µl
2.304 mm
638.940
pl/min
663.544
µl/min
500 µl
3.256 mm
1.27608
nl/min
1.32518
ml/min
1 ml
4.608 mm
2.55582
nl/min
2.65417
ml/min
1 ml
4.699 mm
2.65776
nl/min
2.76004
ml/min
3 ml
8.585 mm
8.87142
nl/min
9.21266
ml/min
5 ml
11.989 mm
17.3013
nl/min
17.9668
ml/min
10 ml
14.427 mm
25.0534
nl/min
26.0170
ml/min
20 ml*
19.050 mm
43.6821
nl/min
45.3622
ml/min
30 ml*
21.590 mm
56.1073
nl/min
58.2653
ml/min
60 ml*
26.594 mm
85.1297
nl/min
88.4040
ml/min
APPENDIX B: LEGATO 950 & 952 OEM MIN/MAX FLOW RATES
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com
* Not available on Legato 952. These pumps work with 0.5
µl to 10 ml syringes