Nominal Minimum/Maximum Flow Rates for Various Syringes.
(Actual Limits will vary depending on syringe manufacturer)
Syringe Size
Syringe ID
Minimum Rate
Units
Maximum Rate
Units
0.5 µl
0.103 mm
0.54000
pl/min
596.496
nl/min
1 µl
0.1457 mm
1.14000
pl/min
1.19358
µl/min
2 µl
0.206 mm
2.28000
pl/min
2.38598
µl/min
5 µl
0.343 mm
6.36000
pl/min
6.61487
µl/min
10 µl
0.485 mm
12.7200
pl/min
13.2256
µl/min
25 µl
0.729 mm
28.7400
pl/min
29.8805
µl/min
50 µl
1.030 mm
57.4200
pl/min
59.6496
µl/min
100 µl
1.457 mm
114.900
pl/min
119.350
µl/min
250 µl
2.304 mm
287.400
pl/min
298.468
µl/min
500 µl
3.256 mm
573.960
pl/min
596.076
µl/min
1 ml
4.608 mm
1.14960
nl/min
1.19387
ml/min
1 ml
4.699 mm
1.19550
nl/min
1.24149
ml/min
3 ml
8.585 mm
3.99042
nl/min
4.14394
ml/min
5 ml
11.989 mm
7.78230
nl/min
8.08163
ml/min
10 ml
14.427 mm
11.2692
nl/min
11.7027
ml/min
LEGATO 958 OEM MIN/MAX FLOW RATES