beautypg.com

JUMO 902030 Screw-in RTD Temperature Probe with Form J Terminal Head Data Sheet User Manual

Technical data

background image

2010-03-04/00311469

Data Sheet 90.2030 (90.2003)

Page 1/4

JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,

36039 Fulda, Germany

Postal address: 36035 Fulda, Germany
Phone:

+49 661 6003-0

Fax:

+49 661 6003-607

e-mail:

[email protected]

Internet:

www.jumo.net

JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM 20 2TT, UK
Phone: +44 1279 635533
Fax:

+44 1279 635262

e-mail:

[email protected]

Internet: www.jumo.co.uk

JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone:

315-697-JUMO
1-800-554-JUMO

Fax:

315-697-5867

e-mail:

[email protected]

Internet: www.jumo.us

Screw-in RTD temperature probe
with terminal head form J

For temperatures from -50 to +400°C
As single or double RTD temperature probes
Available with transmitter
Protection class IP65

Screw-in RTD temperature probes are preferentially used for measuring temperatures in liquids
and gases. A decisive selection criterion is the reliable sealing feature of this installation type
with vacuum and with overpressure. The application areas are, among others, in the air
conditioning technology and refrigeration engineering as well as the HVAC, kiln and apparatus
engineering sector.

The measuring insert is normally fitted with a Pt100 temperature sensor according to DIN EN
60751, Class B in 2-wire circuit; versions with Pt500 or Pt1000 temperature sensors can also
be supplied, as well as 3- and 4-wire circuit connections. A transmitter can be optionally
integrated into the connection head.

Technical Data

Terminal head

Form J, die-cast aluminum, M 16x1.5; IP65, ambient temperature -40 to +100 °C
Caution: reduced ambient temperature range when using transmitters,
Data Sheet 70.7030

Process connection

thread, stainless steel 1.4571

Protection tube

Stainless steel 1.4571, Ø 6 mm and Ø 8 mm

Measuring insert

Pt100 temperature sensor according to DIN EN 60751, Class B, two-wire circuit

Response times

t

0.5

= 5 sec, t

0.9

= 14 sec, in water 0.4 m/sec, 6 mm dia.

Transmitter

Analog transmitter, output 4 to 20 mA, Data sheet 70.7030

Accessories

Sheath, Data Sheet 90.9710 (90.9721)