JULABO ME Heating Immersion Circulator User Manual
Page 47

ME
47
 The parameter flashes, switch by pressing
and
.
A >3-point< calibration is carried out.
The value >TMPVAL< is only indicated
In addition the measured value >CALVAL X< is saved 
during the following step 
 Integer digits flash, set by pressing
(79) +
 Decimal digits flash, set by pressing
(70) +
The first of 3 points is calibrated. 
 
Return to 2. Set working temperature value SETPNT: 120.00 °C
The value is only indicated
 Integer digits flash, set by pressing
(119) +
 Decimal digits flash, set by pressing
(50) +
The second of 3 points is calibrated. 
 
Return to 2. set working temperature value SETPNT: 160.00 °C
The value is only indicated.
 Integer digits flash, set by pressing
(159) +
 Decimal digits flash, set by pressing
(30) +
The 3-point calibration is completed
5. Return to standard operation
 Set >YES< after calibration.
(Standard operation)
- AWC100 Compact Recirculating Cooler (15 pages)
 - CF30 Cryo-Compact Circulators (35 pages)
 - CF31 Cryo-Compact Circulators (76 pages)
 - ED Heating Immersion Circulator (23 pages)
 - ED-5A/B Bath Circulators (24 pages)
 - ED-5A Open Bath Circulators (24 pages)
 - EH Heating Immersion Circulator (27 pages)
 - EH-5 Open Bath Circulator (28 pages)
 - F250 Recirculating Cooler (35 pages)
 - F12-ED Refrigerated / Heating Circulator (26 pages)
 - F12-EH Refrigerated and Heating Circulators (32 pages)
 - F12-MA Refrigerated and Heating Circulators (65 pages)
 - F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath (69 pages)
 - F38-EH Refrigerated and Heating Circulators (30 pages)
 - F25-ME Refrigerated and Heating Circulators (66 pages)
 - F25-HE Refrigerated and Heating Circulators (86 pages)
 - F34-ED Refrigerated / Heating Circulator (25 pages)
 - F25-HL Refrigerated Circulators (94 pages)
 - FL7006 Recirculating Coolers (34 pages)
 - FL2503 Recirculating Coolers (36 pages)
 - FL20006 Recirculating Coolers (32 pages)
 - FL1201 Recirculating Coolers (33 pages)
 - FL300 Recirculating Coolers (48 pages)
 - FK30-SL Calibration Baths (88 pages)
 - FT200 Immersion Coolers (20 pages)
 - FC600_600S Recirculating coolers (37 pages)
 - FC1200S Recirculating cooler (45 pages)
 - F81-HL Ultra-Low Refrigerated Circulator (93 pages)
 - F70-ME Ultra-Low Refrigerated Circulators (64 pages)
 - FP52-SL Ultra-Low Refrigerated Circulators (103 pages)
 - FP51-SL Ultra-Low Refrigerated Circulators (90 pages)
 - FP45-HE Refrigerated and Heating Circulators (79 pages)
 - HL-4 Heating Circulators (89 pages)
 - HE-4 Heating Circulators (78 pages)
 - F81-HL Ultra-Low Refrigerated Circulators (93 pages)
 - FT402 Immersion Coolers (24 pages)
 - ME-4 Heating Circulators (62 pages)
 - ME-31A Visco-Baths (62 pages)
 - MB-17 Open Heating Bath Circulators (52 pages)
 - MB Heating Immersion Circulator (48 pages)
 - Magnum 91 Temperature System (77 pages)
 - MA-4 Heating Circulators (51 pages)
 - MA Heating Immersion Circulator (50 pages)
 - LH 47 Highly Dynamic Temperature Control Systems (76 pages)
 