JULABO ME Heating Immersion Circulator User Manual
Page 39

ME
39
9.5.2. Dynamic
internal
°C
t
APER
NORM
Setpoint
temp. stability
°C
t
APER
NORM
temperature ramp
temp. stability
This parameter affects the temperature sequence only in 
case of internal control. 
Factory setting:
APER (aperiodic)
Possible parameters:
NORM
Allows for reaching the setpoint faster – with
setpoint change or ramp function – but 
overshooting of up to 5 % is possible. 
APER
Ramp function: the increase of temperature 
occurs temporally offset and achieves the 
target temperature without overshooting. 
Setpoint change: The temperature increases at 
the same rate, the target temperature is 
achieved without overshooting. 
With both settings constant temperature is achieved 
after approximately the same time. 
9.5.3.
Control parameters – XPU-, XP-, TN-, TV- EXTERNAL
Setting range: 0.1 ...99.9
Setting range: 3 ...9999
Setting range: 0 ... 999
In most cases the control parameters preset in the factors are adequate 
for achieving an optimum temperature sequence. 
The control parameters allow adjustment to special control processes. 
Setting range: 0.1 ... 99.9
Proportional range >Xpu<
The proportional range Xpu of the cascaded controller is only needed for 
external control. 