background image

Preparations

20

Pt100

RS 232

15

A
M
P

15

A
M
P

V

JULABO MVS

27

25a

25b

26

For applications near ambient
temperature (20 °C), the cooling water
temperature should at least be 5 °C
below the working temperature.

By installing the MVS controller (26)
and a solenoid valve (27), sudden
temperature increases can be rapidly
reversed by tap water flowing through
the built-in cooling coil.
For applications requiring constant
near-ambient temperature, tap water is
thus only used in small amounts.
The circulator provides the necessary
control pulse via the connector (15).

Cooling coil connectors (25a, 25b)

Order No.

Description

9 790 000

MVS controller

8 980 700

Solenoid valve

FD200 Flow-through cooler
Order No. 9 655 825

The FD200 is suitable as cooling device for liquids in closed circuits
(loops).
This unit is integrated in the return flow circuit from the external
system connected to a circulator, and continuously withdraws heat
from the liquid.

NOTE: Immersion coolers are used for countercooling instead of tap
water in applications below room temperature.
Order No.

Description

9 650 820

FT200

9 650 840

FT400

8 970 400

Clamp for cooler probe for stainless steel or

plexiglass

bath

tanks