beautypg.com

JULABO FD200 Flow-Through Cooler User Manual

Page 8

background image

Operator responsibility – Safety instructions

8

2.4. Technical

specifications

FT200

FD200

Temperature range

°C

-20 … +30

10 … +30

Cooling capacity
(medium ethanol)

°C
kW

+20 0 -30
0.25 0.15 0.04

+20 +10
0.22 0.18

Refrigerant

R134a

R134a

Recommanded flow rate

l/min

-----

2 - 3

Freezing protection

°C

-----

10

Immerson probe (Lxdia.)

cm

9x4

-----

Connection tubing (L)

cm

120

-----

Dimensions (WxLxH)

cm

18x27x39

18x27x39

Weight kg

18

16

Ambient temperature

°C

5 ... 35

5 ... 35

Mains power connection

V/Hz

190-253 /50

230/50

Current draw (at 230 V)

A

1.5

1.5

Mains power connection

V/Hz

103-127 / 60

115 / 60

Current draw (at 115 V)

A

3.0

3.0



FT400

FT900

Temperature range

°C

-40 ... +30

-90 ... +30

Cooling capacity
(medium ethanol)

°C
kW

+20 +10 -20 -40
0.45 0.36 0.14 0.03

+20 +10 -40 -80
0.3 0.27 0.2 0.7

Cooling compressor

1-stage

2- stage

Refrigerant

R404A

R404A/R23

Immerson probe (Lxdia.)

cm 12x5

65x1.5

(flexible)

Connection tubing (L)

cm

120

160

Dimensions (WxLxH)

cm

20x30x43

38x55x60

Weight

kg

24

50

Ambient temperature

°C

5 ... 35

5 ... 35

Mains power connection

V/Hz

190-253 /50

190-253 / 50-60

Current draw (at 230 V)

A

4

5.5

Mains power connection

V/Hz

103-127 / 60

103-127 / 60

Current draw (at 115 V)

A

3.5

7.0



Note:
All measurements have been carried out at:
rated voltage and frequency; ambient temperature 20 °C;

This manual is related to the following products: