beautypg.com

4 parameter table – INFICON Cube CDGsci (REST Services) User Manual

Page 9

background image

tira88e1-a (2014-03) CDGsci.cp

9

4 Parameter

Table

Parameter name

ASCII

command

Data

type

Access

right

Comment

Reset

RST

unit8

W

Power On Reset [RST 0] or [q RST]

FilterSettings

FIL

unit8

RW

Filter Settings: 0 = dynamic

1 = fast

2 = slow

3 = Bypass

SP1LevelLow S1L

real32

RW

Setpoint 1 switch on, pressure units

SP2LevelLow S1L

real32

RW

Setpoint 2 switch on, pressure units

SP1LevelHigh

S1H

real32

RW

Setpoint 1 switch off, pressure units

SP2LevelHigh

S2H

real32

RW

Setpoint 2 switch off, pressure units

PerOfAtmSP1 S1P

uint8

RW

Setpoint

1,

percentage of atmosphere (only for 1000Torr

gauges). Not interchangeable with S1[L/H]. Reset S1P to
0 before using S1[L/H].

PerOfAtmSP2 S2P

uint8

RW

Setpoint

2, percentage of atmosphere (for 1000 Torr

gauges only). Not interchangeable with S1[L/H]. Reset
S1P to 0 before using S1[L/H].

ZeroAdjust

ZAD

uint8

W

Zero adjust gauge

ZeroAdjValue

ZAV

sint16

RW

Value after Zero Adjust, unit counts [-2^15…2^15]

DcOutputOffset DOO

sint16

RW

Customer

DC-Output Offset, unit counts [-2^15…2^15]

RemainingZero RZE

sint16

R

Remaining

Zero value, unit counts [-2^16…2^15]

FirmwareRevisionCPU2

SSV

string

R

Software version EtherCAT

ImageRevisionCPU2

AIM

string

R

Software version Cube

FirmwareRevisionCPU1

SWV

unit8

R

Software version CPU1

SwDateYear

SWY

string

R

Year of software version in readable format (string)

SwDateMonthDay SWD

string

R

Month

and day of software version [MMDD]

CalibDate CDA

string

R

Calibration date (xY'MM'DD'HH'MM)

PartNo PAN

string

R

Part number on custom label (max. 20 byte incl. null
terminator)

SerialNumber

SNU

unit32

R

Serial number (unique)

RunHours

RHO

unit16

R

Running hours: 1 = 1h

ExtendedError EXE

uint16

R

Extended

Errors, coded as 16bit uint.

High byte: Bit 0 PT1000 fault

Bit 1 Heater block overtemperature

Bit 2 Electronic overtemperature

Bit 3 Zero adjust error

Low byte: Bit 0 Atm. pressure out of range

Bit 1 Temperature out of range

Bit 4 Cal. mode wrong

Bit 5 Pressure underflow

Bit 6 Pressure overflow

Bit 7 Zero adjust warning

SensPressRange

SPR

unit8

R

Exponent: 0 = E-3

4 = E+1

1 = E-2

5 = E+2

2 = E-1

6 = E+3

3 = E0

SensFSR

SFS

unit8

R

Mantissa: 0 = 1.0

3 = 2.5

1 = 1.1

4 = 5.0

2 = 2.0

5 = 1.4

Help

HLP

string

R

Returns all the available commands

SystemDateTime SDT

string

RW

Set

date and time. Format: DD/MM/CCYY hh:mm:ss

ComportCPU2 COA

string

RW

RS baudrate for CPU2 processor board. Selectable
values are 9600, 19200, 38400, 57600

Ethernet LAN

CLA

string

R

Ethernet LAN on/off

(continued)