CHC-15 CRYSTAL HOLDER & GC-15 GLASS CELL

The figure below shows a INFICON 1” diameter as seen from the front side.

Figure 6.

1" Crystal - as Seen From The Front Side

2.2.2 CRYSTAL PARAMETERS

Polished one-inch diameter crystals that are commonly available for liquid work have the typical values

as listed below.

Type

Frequency

Range (MHz)

Electrode

Material

Resistance

(ohms)

Q Factor

5 MHz

4.976 – 5.020

Gold

~10

120,000

9 MHz

8.976 – 9.036

Gold

~7

55,000

2.2.3 CRYSTAL SURFACE FINISH

Studies have shown that electrode surface roughness can cause large apparent mass loadings due to the

liquid that is trapped within pores at the crystal surface

1

. INFICON’s crystals are optically polished to 50

Å average surface roughness to minimize this effect. Polished crystals are required to obtain good

agreement between theory and measurement during liquid immersion experiments. Polished crystals are

also required to obtain measurements reproducibility from crystal to crystal

2

.

Non-polished crystals (R

a

=1.8 microns) are also available at reduced costs for applications that do not

require the accuracy and reproducibility of the polished crystals.

2.2.4 CRYSTAL ELECTRODE MATERIALS

INFICON’s crystals are available in a variety of electrode materials including Gold, Platinum, Aluminum,

Silver, Titanium, etc. INFICON also offers Gold electrode crystals with an additional SiO2 outer layer to

create a hydrophilic surface needed for some biological applications.

Page 8