Page 10

background image

10

Description

ANALOG IN h
Analog input (input voltage range 0 V to 10.8 V)

Pin assignment:

ANALOG OUT d
Analog outputs (e.g. for logging leakage rate and backing pressure)

Galvanic isolation (max. 60 V DC, 25 V VAC against GND)

Fig. 3

Sample wiring, digital outputs

1

24V supply (output)

2

GND to 24V supply

3

Analog input (0 V to 10.8 V)

4

GND to analog input

Voltage range

0 ... 10 V

Precision

±15 mV offset, additional ±1% from measurement (current
output voltage) as linearity error (at 25 °C)

Resolution

typ. 2.5 mV

Load

> 10 kΩ

PLC_OUT_8

PLC_OUT_7

PLC_OUT_6

PLC_OUT_5

PLC_OUT_4

PLC_OUT_3

PLC_OUT_2

PLC_OUT_1

PLC_OUT_COM

F 0.75 A

8

7

6

5

4

3

2

1

9

F 0.75 A

Trigger 1, inverted

Trigger 2, inverted

Trigger 3, inverted

Trigger 4, inverted

Ready

Error, inverted

CAL request, inverted

Open, inverted