HDA 5500 User Manual

- 11 -

Menu point

Setting

Setting range

Default setting

Calibration of the sensor zero point (CALi.)

The actual measurement is saved as the new zero

point. This is possible in the range +/- 2.5 % of the
display range.

is displayed if a calibration has been

carried out in the permitted range.

is displayed if a calibration has not

been possible.

The function is useful for example, if the output signal always
has an offset which should be displayed as the zero point.

Actual measured value is not to be saved.

YES
no

no

Allocation of analogue output to sensor (ouT.S)

Analogue output allocated to sensor 1

Analogue output allocated to sensor 2

SEn.1
SEn.2
SEn.3
diFF

SEn.1

Analogue output allocated to sensor 3

Analogue output allocated to differential value

(sensor 1 - sensor 2)

Output signal (ouT.M)

4..20 mA output signal

0..10 V output signal

MA
VoLT

VoLT

Software version (VErS)

Display of the software version (for reference only)

To close basic settings (End)

Basis settings menu closes.

Changed settings are saved.

Basic settings can still be altered.

YES
no

no