HYDAC ELECTRONIC GMBH

Part No. 669820

8 Error

message

If an error is detected, a corresponding error message appears that must be
acknowledged by pressing any key.
Possible error messages:

E.10

A data error was detected in the saved settings. This could be due to strong
electromagnetic interference or a component fault.


Action

:

Press and confirm "RES" by pressing "Yes". The factory settings will be
restored for all adjustable parameters and all minimum and maximum values
will be deleted. Enter the data again from the beginning.


E.12

An error was detected in the saved calibration data. This could be due to
strong electromagnetic interference or a component fault.


Action

:

Disconnect then reconnect the supply voltage to the instrument. If the error
persists, the instrument must be returned to the factory for recalibration or
repair.


E.21

A communication error was detected within the instrument. This could be due
to strong electromagnetic interference or a component fault.


Solution

: Press . If the error persists, disconnect then reconnect the supply voltage

to the instrument. If the error still persists, please contact our service
department.