13.0 RAdIATEd EMISSIONS

The use of accessories and cables, other than those specified or sold

by the manufacturer of the equipment as replacement parts for internal
components, may result in increased emissions or decreased immunity of

the equipment.

The T7 Point-of-Care Technology Cart may cause interference when used

adjacent to or stacked with electrical equipment outside the system and
could affect normal operation of said equipment. Do not stack additional
electrical equipment on top of the work surface enclosure.

13.1 RECOMMENdEd SEPARATEd dISTANCES

The T7 is intended for use in the electromagnetic environment specified below in which radiated disturbances are

controlled. The T7 user can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF Communications Equipment and the T7 as recommended below, according to the maximum
output power of the communications equipment.

For transmitters rated at a maximum output power not listed above, the recommended separation distance (D) in meters
(M) can be determined using the equation applicable to the frequency of the transmitter, where (P) is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Max Output Power

(W)

Separation

(M)

Separation

(M)

Separation

(M)

150 kHz to 80 MHz

D = 1.2 (SartP)

81

to

800 MHz

D

=

1.2

(Sart P)

801 MHz

to

2.5 GHz

D

=

1.2

(Sart

P

)

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

Equipment and Systems that are NOT Life-Supporting

Note 1: At 80 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects

and people.