Specifications – Heidolph Incubator 1000 User Manual
Page 24

24
E
SPECIFICATIONS
INCUBATOR 1000
power supply
230/240 V 50/60 Hz / 115 V 50/60 Hz (optional)
temperature range
5 K above ambient (65°C max.)
heating power
300 W
accuracy (temperature
control)
± 2 K up to 50°C
± 4 K above 50°C (air temperature)
temperature distribution
± 2 K up to 50°C
± 3 K above 50°C
time to heat circulating air
abt. 12 minutes from ambient to 50°C
abt. 25 minutes from ambient to 60°C (unobstructed
circulation)
protection class
IP 30
duty cycle
100 %
temperature control
electronic
temperature sensor
Pt 100, installed in heater module
heater
tubular heater with overheat protection
circulation of air
cross current fan, recirculating principle
temperature display
nominal: 2-digit LED display
actual: 2-digit LED display
size (w x d x h)
abt. 170 mm x 375 mm x 97 mm
Hood for Incubator 1000
material
PETG, split-design
size (w x d x h)
abt. 570 mm x 390 mm x 270 mm
volume (air)
abt. 50 dm³
Hood (flat) for Incubator 1000
material
PETG, split-design
size (w x d x h)
abt. 570 mm x 390 mm x 170 mm
volume (air)
abt. 30 dm³