Ie i, Disposal, Malfunction & remedy – Heidolph Multi Reax User Manual
Page 22: Specifications
![background image](/manuals/561969/22/background.png)
I
E
I
22
DISPOSAL
For disposal, please comply with your local or national regulations.
Split by metal, plastic, etc.
Packing material to be treated as described above (material split).
MALFUNCTION & REMEDY
Work on electric, electronic and cryogenic components is limited to qualified personnel.
Master switch on shaker / mixer won't light
1. Check power cord
2. Check circuit breakers
Item won't shake (master switch lighted))
1. Timer run-down
2. Thermal motor circuit breaker triggered by motor overload
Remedy:
Wait about 20 minutes, decrease load applied on shaker plate.
3. Mechanical parts broken (humming motor noise) or electronic failure (no motor noise).
SPECIFICATIONS
Model
Multi Reax
Motion
Vibrating
Rotation speed (rpm)
150
– 2000
Rotation speed setting
Digital
Orbit (mm)
3
Operating mode
Timer or continuous
Input power (W)
33
Weight (kg)
9,8
Dimensions (wxhxd) (mm)
270 x 410 x 172
Standard equipment
Carousel attachment for 26 test tubes, 10 - 16mm
and max. length of 160 mm,
or Carousel attachment for 12 test tubes, 16 -
32mm
and max length of 120 mm
Max. load capacity (kg)
1,5
Supply voltage
230/240V / 50/60 Hz or 115V / 50/60 Hz
Overheating protection
Self-resetting
Protection class (DIN 40050)
IP 30
Permissible ambient temp. (°C)
0
– 50
Permissible relative humidity (%)
80
Designed for continuous use
yes