Heidolph MR Hei-Tec User Manual
Page 22

14
2. Select heating method a. or b.
MR Hei-Standard
a. Heating without temperature sensor
The temperature is measured and regulat-
ed exclusively using the hotplate.
The actual temperature in the stirring con-
tainer is always lower than the preset set
temperature. Factors to consider for tem-
perature differences are the following: heat
transfer resistance of the different stirring
containers, heat transfer resistance of the
quantity to be heated and heat loss due to
radiation.
b. Heating with temperature sensor
NOTICE
Connect temperature sensor
Follow the instructions in chapter
"Installation" - "Accessories (op-
tional)" - "Temperature sensor
Pt 1000".
3. Switch on the device
This manual is related to the following products: