DIGIWEIGH DWP-B2004 User Manual
Page 9
![background image](/manuals/557914/9/background.png)
make matage different.These errors could be eliminated by adjusting
the balance.
In addition,when using the balance for a long time,little errors will
appear because of circumstances such as temperature and humidity.
These errors could also be eliminated by adjusting the balance.
Therefore, when using the balance for the first time or moving it from
a place to another, adjusting is necessary.
2.Standard weight
Choose standard weight ad requirement according to different types:
Types
weight
DWP-B2004
200g
3.Steps of adjusting
DWP-B series balance should warm up for two hours.
For example,nothing on the balance pan,press TAR.the display will
show 0.0000g,then press CAL lightly and quickly, CAL-200 will
appear,at that time put 200g standard weight on the balance pan,
wait for some seconds,adjusting ends when“200.00g’
’appear.
-7-