CRSY102 Operations Manual
Rev G 1/31/2013
Page 26
SECTION 6
– BRIEF TROUBLESHOOTING
Reservoir viewing window does not illuminate (blue)
Is recirculator plugged in?
Is the power switch turned on?
Has the recirculator circuit breaker tripped? If so, reset internal circuit breaker.
Is there water in the reservoir? Add water light will be on.
Water is leaking from unit
Is the recirculator on?
Has a power failure recently occurred?
Are the filter housings tight?
Is the o-ring in the filter housing properly seated?
Is the internal reservoir drain valve open?
Is the recirculator water outlet connected properly to the chamber?
Water is not draining properly from the chamber(s) to the recirculator
Is the recirculator physically lower than the chamber(s)?
Is the chamber drain line (recirculator inlet) kinked? Plugged? Have a rise?
Is the recirculator off?
It is part of the normal operation for
The blue reservoir light to flicker
Water to flow into the reservoir producing a splashing and stirring motion
The red Replace Filters indicator light to flash at startup
The recirculator must be powered on to prevent
returning condensate (water from chamber) from
overflowing inside the recirculator.