beautypg.com

2 specifications, Table 2 specifications – C.B.S. Scientific SG Series User Manual

Page 5

background image

C. B. S. œ Scientific 5

C.B.S. LITE Slab Gels

SECTION 1
General Information

1.1 Introduction

C.B.S. Scientific offers the Slab Gel Systems designed to meet a wide variety of applications.
These systems are designed for separating and characterizing protein and can be adapted for
short sequencing runs with the appropriate accessories. Gel casting can be accomplished using
our convenient and unique Gel Wrap

™, or by using our multiple gel casting chambers (see

Section 6, Slab Gel Kits and Accessories).

Other applications include; mobility shift assays, antibody super shift assays, Rnase protection
assays, DNA footprinting, differential display, DNA paternity testing and forensic analysis.

1.2 Specifications

Constructions:

Buffer chamber, safety cover

Acrylic

Electrodes

Platinum wire .012" diameter

Power cords

FR

Urethane

rated

7500VDC,

200mA, 65

°C

Combs

Teflon

Glass plates

Soda-lime

float

glass

Spacers

PVC

Clamps Polypropylene,

stainless

steel

Safety Certification

EN61010-1-1993

(IEC1010-1)

Table 2
Specifications




Slab Gel Unit:

Shipping
weight

Overall unit
size cm
(l) x (w) x (h)

Distance between
electrodes mm

Recommended
buffer volume
mls

Voltage

Limit

VDC

145mm

7 lb.

24 x 19 x 21.91 250 300

upper

125-200 lower

725

160mm

7 lb.

24 x 19 x 23.33 265 300

upper

125-200 lower

800

170mm

7 lb.

24 x 19 x 24.29 275 300

upper

125-200 lower

850

190mm

7 lb.

24 x 19 x 26.35 295 300

upper

125-200 lower

950

220mm

7 lb.

24 x 19 x 29.21 325 300

upper

125-200 lower

1100

280mm

8 lb.

24 x 19 x 35.24 385 300

upper

125-200 lower

1400

400mm

10 lb.

24 x 19 x 48.07 505 300

upper

125-200 lower

2000