1.5 Safety
Power to the Dual Cool Electrophoresis System is to be supplied by an external DC
voltage power supply that must be ground isolated so that the DC voltage output
floats with respect to ground. For any power supply used, the maximum specified
operating parameters for the units are:
Maximum Limits
300 VDC
30 watts power
190mA current
60°C ambient temperature
Current to the unit, provided from the external power supply, must enter the unit
through the lid assembly, providing a safety interlock to the user. Current to the unit
is broken when the lid is removed.
Do not attempt to use the unit without the
safety lid, and always turn the power supply off before removing the lid, or when
working with the unit in any way. Follow safety precautions specified by the
power supply manufacuturer.
ClearPAGE SDS gels, packages of 10 - Cat. #’s
Acrylamide
%
Gel Size (cm)
W x H
1 well
Cat. #
2D well trough +
1 marker lane
Cat. #
12 well
Cat. #
17 well
Cat. #
8%
10 x 8
-
-
BK00812-10
BK00817-10
10 x10
-
-
FK00812-10
FK00817-10
10%
10 x 8
-
-
BK01012-10
BK01017-10
10 x10
FK01001-10*
-
FK01012-10
FK01017-10
12%
10 x 8
-
-
BK01212-10
BK01217-10
10 x10
FK01201-10*
-
FK01212-10
FK01217-10
16%
10 x 8
-
-
BK01612-10
BK01617-10
10 x10
-
-
FK01612-10
FK01617-10
4 - 8%
10 x 8
-
-
BK40812-10
BK40817-10
10 x10
-
-
FK40812-10
FK40817-10
4 - 12%
10 x 8
-
-
BK41212-10
BK41217-10
10 x10
-
-
FK41212-10
FK41217-10
4 - 20%
10 x 8
-
-
BK42012-10
BK42017-10
10 x10
FK42001-10* FK42002-10*
FK42012-10
FK42017-10
1.4 ClearPAGE™ Precast Gels Selection Chart*
*See page 21 forDNA/Native Precast Gel Selection Chart