beautypg.com

C.B.S. Scientific DGGEK-2401 User Manual

Page 24

background image

24

Cipher DGGE Instructions 2/28/14

SECTION 4

Gel Running Preparations

4 1 Preparation of the Cassettes

1. After polymerization rinse gel plate assembly with D.I. water to remove

excess acrylamide or denaturants from plate exterior.

2. Remove comb and quickly transfer gel

sandwich to cassette by removing all #2

clamps.

LEAVE GEL WRAP IN PLACE

The Gel Wrap acts as a barrier and

prevents perpendicular electrical fields

from interfering with outside lanes. Re-

clamp to cassette using #1 clamps/4

per side.

Dislodge bottom portion of

Gel Wrap from the plates to allow buffer

contact with the gel. See figure 4-1a.

NOTE: If running only 1 gel on a dual

cassette, attach conversion plate with

clamps to second side of cassette.

3. Immerse single (fig. 4-1b) or dual cassette(s) (fig. 4-1c) into pre-heated buffer

tank.

Fig 4-1a

Fig 4-1b

Fig 4-1c

This manual is related to the following products: