C.B.S. Scientific DGGEK-2401 User Manual
Page 17

17
www.cbsscientific.com
3. Place the gasketed plate on the lab
bench with the tubing side up, and
extend the bottom of the plate over the
edge of the bench, approximately ¾ of
an inch. Place the spacers along side
the inside edges of the gasket. Be sure
the rounded corner end of each spacer
is facing the outside bottom of the
plate, following the radius of the glass.
4. Place the notched plate on top of the
bottom assembly, starting from the
bottom edge and gently easing the
plate down. Verify the gasket is smooth
around theedges and then clamp along
the bottom.
5. Lift the assembly and stand it on the
base of the clamps. For leveling, push
glass plate assembly down until it
stops against clamp body. Clamp the
sides of the assembly with additional
casting clamps on either side. As each
clamp is attached, be sure the gasket
is aligned between the plates forming a
seal.