beautypg.com

2 specifications, 3 safety – C.B.S. Scientific GCMGU-602T User Manual

Page 5

background image

5

www.cbsscientific.com

SECTION 1

General Information
1.1 Introduction

The Horizontal Midi-Gel systems, (GC)MGU-402T, (GC)MGU-502T, and (GC)GCMGU-
602T, provide a compact format to rapidly separate and identify DNA and RNA frag-
ments, PCR products and synthetic oligonucleotides, while using minimal quantities of
expensive reagents. The tapered baffle system allows gel casting directly onto the unit
or removable trays without having to seal with agarose plugs or tape. The gel bed and
gel tray are UV transparent. Combs are offered in several thicknesses and formats to
accommodate a wide variety of experimental applications.

1.2

Specifications

Constructions

Buffer chamber, lid

Acrylic

Electrodes

Platinum wire .012” diameter

Power cords

FR Silicone rated 7500VDC, 200mA, 65ºC

Tapered baffles

Polycarbonate

Combs

Polycarbonate, Anodized Aluminum or Teflon

Comb backing

Acrylic

UVT Gel Tray (optional)

U-V transparent acrylic

Safety Certification

EN61010-1-1993 (IEC1010-1)

*Calculated using gel thickness of 0.5cm

1.3

Safety

Power to the Midi-Horizontal units is to be supplied by an external DC voltage power
supply that must be ground isolated so that the DC voltage output floats with respect to
ground. For any power supply used, the maximum specified operating parameters for
the units are:

Maximum Limits

250VDC voltage

15 watts power

60 mA current

50*C ambient temperature

Current to the unit, provided from the external power supply, must enter the unit through
the lid assembly, providing a safety interlock to the user. Current to the unit is broken
when the lid is removed. Do not attempt to use the unit without the safety lid. Always
turn the power supply off before removing the lid, or when working with the unit in any
way. Follow safety precautions specified by the power supply manufacturer.

Model #

(GC)MGU-402T (GC)MGU-502T

(GC)MGU-602T

Shipping weight

4 lb.

6 lb.

7 lb.

Overall Size W x L x H (cm)

13 x 22.5 x 14

16.7 x 21.5 x 14 16.7 x 27.5 x 14

Gel Bed Size W x L (cm)

10.5 x 11

14 x 10

14 x 16

Recommended Buffer Volume

550ml

700ml

750ml

Agarose Solution Required*

50ml

60ml

100ml

# of Slots in Gel Tray

2

2

4

Pathlength/Slot (cm)

4.0

3.5

3.0

Voltage limit

250VDC

250VDC

250VDC

This manual is related to the following products: