Quick Calibration (CALIBR):

Allows the user to perform a quick calibration of the zero and span.

NOTE:

Both zero pressure and full-scale pressure must be applied to the
gauge in order to complete the quick calibration process.

Step 1: Press the On/Off button to enter the main setup menu.
Step 2: Press (Z) to cycle through the menu options until AdVSET is

displayed.

Step 3: Press (S) to enter the Advanced Settings menu.

Step 4: Press (Z) to cycle through the options until CALIBR is displayed.
Step 5: Press (S) to access the CALIBR mode. The gauge will display the

word ZERO, prompting the user to perform the zero pressure quick
calibration.

Step 6: Ensure no pressure is applied to the gauge. Press (Z) to zero the gauge.

The gauge will then display the word SPAN, prompting the user to
perform the full-scale pressure quick calibration.

Step 7: Apply full jack pressure to the gauge and press the (S) button to

complete the quick calibration

and return to advanced setup menu.

Step 9: To exit the advanced setup menu, press (Z) until EXIT is displayed and

press (S) to exit to the main setup menu.