Quick Calibration (CALIBR):
Allows the user to perform a quick calibration of the zero and span.
NOTE:
Both zero pressure and full-scale pressure must be applied to the
gauge in order to complete the quick calibration process.
Step 1: Press the On/Off button to enter the main setup menu.
Step 2: Press (Z) to cycle through the menu options until AdVSET is
displayed.
Step 3: Press (S) to enter the Advanced Settings menu.
Step 4: Press (Z) to cycle through the options until CALIBR is displayed.
Step 5: Press (S) to access the CALIBR mode. The gauge will display the
word ZERO, prompting the user to perform the zero pressure quick
calibration.
Step 6: Ensure no pressure is applied to the gauge. Press (Z) to zero the gauge.
The gauge will then display the word SPAN, prompting the user to
perform the full-scale pressure quick calibration.
Step 7: Apply full jack pressure to the gauge and press the (S) button to
complete the quick calibration
and return to advanced setup menu.
Step 9: To exit the advanced setup menu, press (Z) until EXIT is displayed and
press (S) to exit to the main setup menu.