beautypg.com

1 galaxy o2 sensor setup, 1 delivery package, Galaxy o2 sensor setup — instructions – Eppendorf Galaxy CO2 Sensor User Manual

Page 3: Galaxy o, Sens or se tup

background image

Galaxy O2 Sensor Setup — Instructions

3

1

Galaxy O

2

sensor setup

1

Galaxy O

2

sensor setup

The O

2

sensor used on all New Brunswick Galaxy incubators equipped with the O

2

option is a

sensitive, self-powered, electrochemical cell, with a typical life span of one to two years

dependent on ambient atmospheric levels.
To ensure the maximum shelf life of the O

2

sensor, this incubator is equipped with a placeholder

O

2

sensor. Use the following instructions to replace the supplied O

2

sensor:

During initial setup of new incubators.

After the sensor shelf life has expired.

Sensor is unusable.

1.1 Delivery package

1 O

2

sensor

2 O

2

sensor tool

1

2

1

Galaxy O

2

sens

or

se

tup