6
Technical data
6.1
Test conditions
Test conditions and test evaluation in accordance with ISO 8655, part 6. Test with
analytical balance with evaporation protection, approved by the board of weights and
measures.
Measurement errors do not apply to autoclaved Mastertips.
When using less than 2 μL for the test, the sample liquid and the experience of the user
have a critical influence on accuracy and precision. Therefore no limits are set for this
range.
• Number of determinations: 10
• Use of water in accordance with ISO 3696
• Test with a Mastertip
• Test at 20 °C – 27 °C ±0.5 °C
• Dispensing on the inner wall of the vessel
Technical specifications are subject to change.
Volume
Systematic error
Random error
2 μL
±6.0%
≤4.0%
3 μL
±5.0%
≤3.0%
5 μL
±4.0%
≤2.0%
10 μL
±3.0%
≤1.5%
20 μL
±2.0%
≤0.8%