1. Please note the information in the document "Decontamination certificate for product

returns".
You can find it as a PDF file on our webpage www.eppendorf.com.

2. Enter the serial number of the Pipette in the decontamination certificate.
3. With the shipment please include the completed decontamination certificate for

product returns with the Pipette.

4. Send the pipette to Eppendorf AG or an Eppendorf AG service partner. The address

details of Eppendorf AG can be found on the last page of this operating manual. The list
of service partners can be found at www.eppendorf.com/worldwide.

6

Technical data

6.1

Test conditions

Test conditions and test evaluation in accordance with ISO 8655, part 6. Test with
analytical balance with evaporation protection, approved by the board of weights and
measures.

Measurement errors do not apply to autoclaved Mastertips.

When using less than 2 μL for the test, the sample liquid and the experience of the user
have a critical influence on accuracy and precision. Therefore no limits are set for this
range.

• Number of determinations: 10
• Use of water in accordance with ISO 3696
• Test with a Mastertip
• Test at 20 °C – 27 °C ±0.5 °C
• Dispensing on the inner wall of the vessel

Technical specifications are subject to change.

Volume

Systematic error

Random error

2 μL

±6.0%

≤4.0%

3 μL

±5.0%

≤3.0%

5 μL

±4.0%

≤2.0%

10 μL

±3.0%

≤1.5%

20 μL

±2.0%

≤0.8%