4 electrical connections, Electrical connections – Eppendorf 2150 Digital Platform Shaker User Manual
Page 18

Installation
New Brunswick Digital Platform Shaker Innova® 2100/2150
English (EN)
18
The required surface area required for operation is:
4.4
Electrical connections
Before making electrical connections, be sure to follow these instructions:
1. Check the voltage selector in the Universal Power Module at the rear of the unit to ensure that it is set to
the appropriate voltage.
2. Remove the CAUTION label from the universal power module.
3. Make sure the plastic straps have been removed from the upper bearing housing.
4. Make sure the power switch on the back of the unit is set in the OFF position.
ONLY THEN:
5. Connect the power cord to the universal power module and to a grounded electrical outlet.
Model
Width
Depth
Innova 2100 without capacity upgrade
56 cm (22 in)
61 cm (24 in)
Innova 2150 and 2100 with capacity upgrade
68.5 cm (27 in)
61 cm (24 in)
WARNING!
Do not position the equipment so that it is difficult to operate the disconnecting device
(main plug).
For service reasons and for cases of emergencies, sufficient space must be provided for
easy access to the main power cord/power switch (located at the back of the unit).
WARNING! Risk of electric shock and/or damage to device!
Use a grounded power supply.
CAUTION!
A detachable main power cord has been provided with your unit.
Only use the cord provided with the unit.