6 operation, 1 getting started, 1 plug in – Eppendorf New Brunswick -86 °C Freezers Innova U360 User Manual
Page 18: 2 turning the freezer on/off, Operation 6.1, Getting started 6.1.1, Plug in, Turning the freezer on/off, Op eration
Innova® U360 -86 °C Freezers — Operating manual
18
6
Operation
6
Operation
6.1
Getting started
6.1.1
Plug in
Once you have verified that the mains/power supply matches the electrical requirements of the
freezer, connect the product to the mains/power supply using the mains/power cord provided.
6.1.2
Turning the freezer On/Off
The ON/OFF circuit breaker is located within the lockable panel at the bottom left-hand corner of
the freezer.
To remove the lockable panel and turn the circuit breaker and battery switch On/Off:
1. Insert and turn the key (provided) one quarter turn to the right.
2. Remove the panel.
WARNING!
Risk of personal injury
BEFORE connecting the freezer to the mains/electrical supply, make sure that the mains/
power supply matches the requirements of the equipment. Check the specification plate
(located on the side of the freezer) for the electrical requirements. The equipment should be
connected to an earthed/grounded outlet socket.
WARNING!
Risk of personal injury
If the freezer’s voltage rating does not match your mains/electrical supply, or if the plug on the
mains/power cord does not fit the outlet, do not plug the freezer in.
Contact your laboratory supervisor, safety officer, or qualified service or electrical engineer.
NOTICE!
Risk of material damage
Some freezers are supplied with more than one removable mains/power cord. Utilize the cord
that matches your power receptacle. Check the voltage rating plate on the side of the freezer,
to confirm that the freezer is compatible with your laboratory mains/power supply.
WARNING!
Risk of personal injury
The On/Off circuit breaker and battery switch are fitted with IP65 plastic covers, to prevent a
possible source of ignition. These covers must not be removed. If one of the covers needs to
be replaced, the replacement must be performed by a qualified and authorized person.
Failure to observe this safety warning will invalidate the warranty and could result in a
dangerous situation in the event of a failure.
Hint!
The key can be removed to prevent access.
6
Op
eration