beautypg.com

2 packing list verification, 3 product versions, 1 introduction – Eppendorf New Brunswick -86 °C Freezers Innova U360 User Manual

Page 11: 4 features, Packing list verification, Product versions 3.3.1, Introduction, Features, Product descriptio n

background image

11

Innova® U360 -86 °C Freezers — Operating manual

3.2.2

Packing list verification

Unpack your order, saving the packing materials for possible future use. Save the operating
manual for instruction and reference. Verify against your New Brunswick packing list that you
have received the correct materials, and that nothing is missing. If any part of your order was
damaged during shipping, is missing, or fails to operate, fill out the "Customer Feedback" form,
available online at www.nbsc.com/CustomerFeedback.aspx.

3.3

Product versions

3.3.1

Introduction

This manual provides the user with the necessary information for installation and operation of
New Brunswick's Innova

®

ultra-low temperature freezer with vacuum insulation panels. It also

provides some preliminary user maintenance information.
This manual covers the U360 freezer models:

3.4

Features

The freezers are manufactured using high quality steel and electronics for long life. The cabinets
are insulated with a combination of foamed-in-place polyurethane foam and vacuum insulation
panels. This combination ensures superior insulation properties while maintaining an extra large
internal capacity for the footprint of the freezer. The Innova freezers have many features designed
to provide ease of use and maintanence, security and reliability to your ultra-low temperature
storage needs. They are built to stringent regulatory requirements for safety and environmental
friendliness and disposability, and they are CE and UL certified.
Features include:

Designed to be exceptionally energy efficient and eco-friendly.

A large ergonomically located backlit LCD display provides operational information about the
freezer including temperature and setpoint info, on-screen messaging, programmable alarms,
data graphing, and diagnostics.

Vacuum insulation panel technology reduces the cabinet wall thickness while maintaining
superb insulation characteristics to provide the largest internal volume per floor space used.

Freezers are totally free of CFCs (Chlorofluorocarbons) and HCFCs
(Hydrochlorofluorocarbons). They use industrial grade, commercially available HFCs
(Hydrofluorocarbons) as refrigerants.

Temperature alarms and door opening data are automatically stored for up to 30 days for
on-screen graphing. Data is downloadable to PC with optional communications port for
validation requirements.

Upright models utilize gasketed and insulated inner doors to reduce cold air loss and recovery
times, reduce internal temperature fluctuations and conserve energy.

NOTICE!

Risk of material damage

Vacuum insulation panels are used in the construction of these freezers. Inspect the cabinet
panels for punctures or other damage that compromises the integrity of the product.

These panels are mounted in the cavity against the steel outer wall of the freezer. Any drilling
or puncture to the outer wall could release the vacuum from the panel, resulting in impaired
freezer performance.

Any unauthorized punctures or other damage deliberately made to the cabinet walls will
invalidate the warranty.

Model (230 V, 50 Hz)

Capacity

U360

360 liters (12.7 cubic feet)

Model (120 V, 60 Hz)

Capacity

U360

360 liters (12.7 cubic feet)

3

Product descriptio

n