9.4 Application parameters

9.4.1

Tempering

9.4.2

Mixing

9.5 Interface

Temperature control

range

Minimum: 4 °C above ambient temperature
Maximum:100 °C
Temperature setting 1 °C – 100 °C, adjustable in steps of 1 °C

Temperature accuracy

Set temperature
20 °C – 45 °C

Set temperature
< 20 °C or > 45 °C

ThermoMixer F1.5

±0.5 °C

±0.5 °C

ThermoMixer FP

±1.0 °C

±4.0 °C

Temperature homogeneity In the range of 20 °C – 45 °C max. ±0.5 °C

for all positions of the thermoblock

Heating rate

ThermoMixer F1.5: 11 °C/min
ThermoMixer FP: 18 °C/min
The change of temperature in filled tubes is slower.

Mixing frequency can be set in increments of 50 rpm

ThermoMixer F1.5

300 rpm – 1 500 rpm

ThermoMixer FP

300 rpm – 2 000 rpm

USB interface

For Eppendorf service only.