4 application parameters, 1 tempering, 2 mixing – Eppendorf ThermoMixer FP User Manual
Page 27: 5 interface, 1 tempering 9.4.2 mixing, Eppendorf thermomixer f1.5/fp — operating manual
Eppendorf ThermoMixer F1.5/FP — Operating manual
27
9.4 Application parameters
9.4.1
Tempering
9.4.2
Mixing
9.5 Interface
Temperature control
range
Minimum: 4 °C above ambient temperature
Maximum:100 °C
Temperature setting 1 °C – 100 °C, adjustable in steps of 1 °C
Temperature accuracy
Set temperature
20 °C – 45 °C
Set temperature
< 20 °C or > 45 °C
ThermoMixer F1.5
±0.5 °C
±0.5 °C
ThermoMixer FP
±1.0 °C
±4.0 °C
Temperature homogeneity In the range of 20 °C – 45 °C max. ±0.5 °C
for all positions of the thermoblock
Heating rate
ThermoMixer F1.5: 11 °C/min
ThermoMixer FP: 18 °C/min
The change of temperature in filled tubes is slower.
Mixing frequency can be set in increments of 50 rpm
ThermoMixer F1.5
300 rpm – 1 500 rpm
ThermoMixer FP
300 rpm – 2 000 rpm
USB interface
For Eppendorf service only.
EN
Operating manual