5.5 Temperature control

The ThermoMixer F1.5/ThermoMixer FP can be used for temperature control in a range of 4 °C

above the ambient temperature to 100 °C.

5.5.1

Temperature control without mixing process

1. To switch off the mixing function, use the speed arrow keys to select the 0 rpm setting

(» below 300 rpm or º above 1 500 rpm or 2 000 rpm).

2. Using the temp arrow keys set the temperature.

The device immediately starts to perform temperature control.

When the set temperature is not changed, the actual temperature flashes on the display

and the device does not perform temperature control.

3. To manually start the temperature control procedure, press the start/stop key.

The  symbol flashes on the display.

The display shows the actual temperature/set temperature.

NOTICE!

NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to

a warmer environment.

After installing the device, wait at least for 3 h. Only then connect the device to the

mains.

Hint!

When the actual temperature flashes on the display, the device is not in

temperature control mode operation.

As soon as the set temperature is changed with the temp arrow keys, the device

begins to perform temperature control.

When the set temperature has been reached, the display only shows one value.