Eppendorf ThermoMixer F1.5/FP — Operating manual
13
3.5 Warning signs on the device
4
Installation
4 Installation
4.1 Selecting the location
Select the location for the device according to the following criteria:
•
Suitable power connection as per the name plate.
•
Minimum distance to other devices and walls: 10 cm (3.9 in).
•
Bench with a horizontal and even work surface which is designed to support the weight of the
device.
•
Surrounding area must be well ventilated.
•
Location must be protected against direct sunlight.
4.2 Installing the instrument
1. Place the ThermoMixer F1.5/ThermoMixer FP onto a suitable work surface.
Position the device in such a way that the ventilation slots on the underside of the device are
not obstructed.
2. Connect the power cable to the power connection socket of the device and the power supply.
Representation Meaning
Location
Risk of burns from hot surfaces.
Upper device side
Observe the operating manual.
Rear of the device
Caution! Risk of injury from moving parts.
Observe the operating manual.
On the
thermoblock
Hint!
The mains/power switch and cutting unit of the mains/power line must be easily
accessible during operation (e.g, residual current circuit breaker).
WARNING!
WARNING! Risk from incorrect supply voltage
Only connect the device to voltage sources which correspond to the electrical
requirements on the name plate.
Only use sockets with a protective earth (PE) conductor and suitable power cable.