4 application parameters, 1 temperature control, 2 mix – Eppendorf ThermoMixer FP User Manual
Page 31: 5 interface, Application parameters 9.4.1, Temperature control, Interface
Technical data
Eppendorf ThermoMixer
®
F0.5/F1.5/F2.0/FP
English (EN)
31
9.4
Application parameters
9.4.1
Temperature control
9.4.2
Mix
9.5
Interface
Temperature control range
Minimum: 4
°C above ambient temperature
Maximum: 100
°C
Temperature setting 1
°C
–
100
°C,
adjustable in steps of 1
°C
Temperature accuracy
Set temperature
20
°C
–
45
°C
Set temperature
<
20
°C or >
45
°C
Eppendorf ThermoMixer
F0.5
Eppendorf ThermoMixer
F1.5
Eppendorf ThermoMixer
F2.0
Eppendorf ThermoMixer
FP
±0.5
°C
±0.5
°C
±0.5
°C
±1.0
°C
±0.5
°C
±0.5
°C
±0.5
°C
±4.0
°C
Temperature homogeneity
in relation to all positions of the
thermoblock
Set temperature
20
°C
–
45
°C
Set temperature
<
20
°C or >
45
°C
Eppendorf ThermoMixer
F0.5/F1.5/F2.0/FP ±0.5
°C
±1.5
°C
Heating rate
Eppendorf ThermoMixer
F0.5:
15
°C/min
Eppendorf ThermoMixer
F1.5:
11
°C/min
Eppendorf ThermoMixer
F2.0:
13
°C/min
Eppendorf ThermoMixer
FP:
18
°C/min
The change of temperature in filled tubes is
slower.
Mixing frequency can be set in increments of 50
rpm
Eppendorf ThermoMixer
F0.5
300
rpm
–
2
000
rpm
Eppendorf ThermoMixer
F1.5
300
rpm
–
1
500
rpm
Eppendorf ThermoMixer
F2.0
300
rpm
–
1
500
rpm
Eppendorf ThermoMixer
FP
300
rpm
–
2
000
rpm
USB interface
For Eppendorf service only.