Technical data

Eppendorf ThermoMixer

®

F0.5/F1.5/F2.0/FP

English (EN)

31

9.4

Application parameters

9.4.1

Temperature control

9.4.2

Mix

9.5

Interface

Temperature control range

Minimum: 4

°C above ambient temperature

Maximum: 100

°C
Temperature setting 1

°C

100

°C,

adjustable in steps of 1

°C

Temperature accuracy

Set temperature
20

°C

45

°C

Set temperature
<

20

°C or >

45

°C

Eppendorf ThermoMixer

F0.5
Eppendorf ThermoMixer

F1.5
Eppendorf ThermoMixer

F2.0
Eppendorf ThermoMixer

FP

±0.5

°C
±0.5

°C
±0.5

°C
±1.0

°C

±0.5

°C
±0.5

°C
±0.5

°C
±4.0

°C

Temperature homogeneity
in relation to all positions of the
thermoblock

Set temperature
20

°C

45

°C

Set temperature
<

20

°C or >

45

°C

Eppendorf ThermoMixer

F0.5/F1.5/F2.0/FP ±0.5

°C

±1.5

°C

Heating rate

Eppendorf ThermoMixer

F0.5:

15

°C/min

Eppendorf ThermoMixer

F1.5:

11

°C/min

Eppendorf ThermoMixer

F2.0:

13

°C/min

Eppendorf ThermoMixer

FP:

18

°C/min

The change of temperature in filled tubes is
slower.

Mixing frequency can be set in increments of 50

rpm

Eppendorf ThermoMixer

F0.5

300

rpm

2

000

rpm

Eppendorf ThermoMixer

F1.5

300

rpm

1

500

rpm

Eppendorf ThermoMixer

F2.0

300

rpm

1

500

rpm

Eppendorf ThermoMixer

FP

300

rpm

2

000

rpm

USB interface

For Eppendorf service only.