2 inserting tubes and plates, Inserting tubes and plates – Eppendorf ThermoMixer FP User Manual
Page 18
18
Operation
Eppendorf ThermoMixer
®
F0.5/F1.5/F2.0/FP
English (EN)
5.2
Inserting tubes and plates
Inserting the plate
Insert the plate with the back edge first. Then press it down at the front.
Inserting tubes
Insert the tubes completely into the bores of the thermoblock.
Only Eppendorf ThermoMixer
FP:
1
Height
sensor
NOTICE! Damage to plates due to too high temperatures.
Polystyrene microplates melt at temperatures above 70
°C.
Polypropylene deepwell plates deform at temperatures above 80
°C.
Deformed
plates can become detached from the thermoblock.
Only heat microplates up to 70
°C.
If you are heating deepwell plates above 80
°C, do not exceed the mixing
frequency of 1000
rpm.
NOTICE! Material change of consumables due to extreme temperatures.
Extreme temperatures (e.g., during refrigeration or autoclaving) affect
consumables material. The mechanical strength, dimensions and shape of the
consumable will change.
Use consumables that are suitable for the selected temperature range or
selected procedure.
The height sensor of the Eppendorf ThermoMixer
FP
automatically
differentiates
between deepwell plates and microplates.
When inserting microplates, make sure that the height sensor is not covered.
Take care that the height sensor does not get contaminated.
To ensure uniform temperature control,
place the lid on the thermoblock.
1