3 interrupting or stopping the program, 1 interrupting the program, 2 continuing the program – Eppendorf Mastercycler pro User Manual
Page 53: 3 stopping the program, 4 using protocols, Interrupting or stopping the program 7.3.1, Interrupting the program, Continuing the program, Stopping the program, Using protocols

PCR run
Mastercycler pro
English (EN)
53
7.3
Interrupting or stopping the program
You can temporarily interrupt the program run in the status view, for instance, to add reagent, or
completely stop the program.
7.3.1
Interrupting the program
1. Open status view (see Starting the program on p. 51).
2. Press the Stop key.
The status view shows
Paused
.
Observe the safety precautions for opening the heated lid (see Opening the heated lid on p. 28).
7.3.2
Continuing the program
1. If the heated lid is opened, draw the heated lid forwards over the samples and close the lid.
2. Press the
Resume
function key.
The status view displays
running
and the program run is continued.
7.3.3
Stopping the program
Press the
Abort
function key.
The status
Idle
is displayed. The program has been aborted and cannot be continued.
7.4
Using protocols
The Mastercycler pro creates a protocol for each program run where all executed program steps and any
interruptions and errors are recorded.
After running a PCR program, you can view the associated protocol on the control panel, save it on an MMC
or a USB memory or send it to a printer.
Prerequisites
The cycler is idling.
1. Mark the desired cycler node in the navigation tree.
2. Press the function key
Protocol
.
The protocol is overwritten the next time a program is started on the Mastercycler pro.