12. Press the
OK
function key to accept the settings.
The program view appears.
13. Press the
>>
function key and then
Save
.
The changes are saved.
14. Press the
Exit
function key to exit the editor.
6.3.3
Deleting individual program steps
1. Use the arrow keys to mark the desired program step.
The marked program step is highlighted in blue.
2. Press the Delete key.
3. Press the
Save
function key to save the change.
Press the
Exit
function key to exit the editor.
Temp
Temperature for the program step.
Equivalent to directly editing in the program view.
Temp Inc./Dec.
For changing the temperature to the next cycle.
Values ranging from -10.0 °C to +10.0 °C in increments of 0.1 °C are possible.
You can switch between positive and negative values by pressing the
+/-
function key.
Gradient
Column temperature gradient in the thermoblock.
Enter the temperature range into the
from
(left-hand side of block) and
to
(right-hand
side of block) fields. The
Temp
and
Temp Incr.
fields are deactivated.
To display the temperatures in the individual columns, press the
ShowGrad
function
key. The temperatures displayed there are interpolated approximation values.
Maximum gradient range and course depend on the type of block used. To use the
extended gradient spread on devices with silver block (Mastercycler pro S and
Mastercycler ep gradient S), define the appropriate cycler type in the program header
in advance(see p. 42).
TIME
Hold time for the program step.
Equivalent to directly editing in the program view.
Time Inc./Dec.
For changing the hold time to the next cycle.
Values ranging from -01:00 to +01:00 in increments of 1 s are possible.
You can switch between positive and negative values by pressing the
+/-
function key.
Ramp
Temperature control speed of the thermoblock.
Values ranging from 1 % to 75 % and also 100 % are possible.