Product description
Mastercycler pro
English (EN)

10

Abb. 2-2: Terminal block

Fig. 2-2:

Terminal block

Abb. 2-3: Control panel

Fig. 2-3:

Control panel

Only devices meeting the IEC 950/EN 60950 (UL 1950) standards may be connected to the interfaces of the
Mastercycler pro and the control panel.

1

Parallel interface
Connection for printer.

2

Serial interface (RS 232)
Connection for Eppendorf temperature validation
system.

3

Control panel connection

4

CAN out
Data outlet for connecting several
Mastercycler pro or Mastercycler ep.

5

CAN in
Data inlet for connecting several
Mastercycler pro or Mastercycler ep.

6

Term switch
Termination of CAN bus when networking several
Mastercycler pro or Mastercycler ep.
Lower position: ON. Upper position: OFF.

7

Address dial
Setting of device address when networking
several Mastercycler pro or Mastercycler ep.

1

Display

2

USB port 1
For mouse, printer or storage medium.

3

USB port 2
For mouse, printer or storage medium.

4

MMC slot
For data exchange via a MultiMediaCard (MMC).

Printer

RS 232

Control Panel

CAN_out CAN_in Term Address

1

2

3

4

5

6

7

1
2

3

4