3 features, 4 rotor fa-45-18-11, Features – Eppendorf 5418 R Centrifuge User Manual
Page 9: Rotor fa-45-18-11
Product description
Centrifuge 5418 / 5418 R
English (EN)
9
2.3
Features
The compact and easy-to-use Centrifuge 5418 / 5418 R has a capacity of 18 x 2 ml and reaches a maximum
of 16,873 x g / 14,000 rpm. The microcentrifuge is equipped with an aerosol-tight standard rotor for
centrifugation of the following tubes:
• Micro test tubes (0.2 to 2.0 ml)
• Microtainers (0.6 ml)
• Spin columns (1.5/2.0 ml)
The Centrifuge 5418 R has an additional temperature control function for centrifugation between 0°C and
+40°C. The
fast temp function can be used to start a temperature control run without samples to adjust the
rotor chamber incl. rotor and adapters quickly to the set target temperature.
2.4
Rotor FA-45-18-11
Before using tubes, observe the manufacturer's recommended specifications on resistance to
centrifugation (max. g-force).
(1) Maximum load per rotor bore for adapter + tube + contents.
(2) According to DIN 58 970 (device version: 230 V, 50 to 60 Hz).
(3) Aerosol tightness tested and certified by the Centre of Emergency Preparedness and Response, Health
Protection Agency, Porton Down (UK) (see certificates at the end of this operating manual).
Max. capacity
Max.
g-force
(rcf)
or
speed
(rpm)
without
adapter
Max. load
per rotor
bore
(1)
Notes
Acceleration/deceleration
time
(2)
Rotor
FA-45-18-11
18 micro test tubes of 1.5/
2.0
ml each or spin columns.
With adapters:
• 0.2 mL PCR tubes
• 0.4 ml/0.5 ml micro test
tubes
• 0.6 mL Microtainers
16,873 x g /
14,000 rpm
3.75 g
• Aerosol-tight
(3)
rotor lid
(aluminum).
5418: 16 s / 18 s
5418 R: 13 s / 13 s
Rotor code:
All Eppendorf
®
rotors are identified
using a simple, alphanumeric
format that represents the technical
specifications in a uniform series of
letters and numbers.
A
45
30
11
Aerosol-tight
version
F
Fixed-angle
rotor
Angle of
bores
Ø of
bores (mm)
Number of
bores
A
Swing-bucket
rotor
4
81
Ø of
buckets (mm)
Number of
buckets