Safety
Centrifuge 5418 / 5418 R
English (EN)

16

3.5.5

Aerosol-tight centrifugation

3.6

Safety instructions located on the device

WARNING! Risk to health due to limited aerosol tightness with incorrect rotor/rotor lid
combination.
Aerosol-tight centrifugation is guaranteed only if the rotors and rotor lids intended for this
purpose are used. For fixed-angle rotors the labeling always begins with

FA, swing-bucket

rotors are labeled with

AT (aerosol tight).

The aerosol-tight rotors and rotor lids of this centrifuge are additionally marked with a red
ring on the rotor and a red rotor lid screw.

For aerosol-tight centrifugation, always simultaneously use rotors and rotor lids which are
marked as aerosol-tight in the centrifuge intended for the corresponding purpose. The
details specifying in which centrifuge you may use the aerosol-tight rotors and rotor lids
can be found on the rotor and, beginning from production date of October 2003, on the
upper side of the rotor lid.

Only use aerosol-tight rotor lids in combination with rotors which are marked on the rotor
lid.

WARNING! Health hazard from limited aerosol-tightness due to incorrect use.
Autoclaving, mechanical stresses and contamination by chemicals or other aggressive
solvents can impair the aerosol-tightness of the rotors and rotor lid.

Check the integrity of the seals of the aerosol-tight rotor lids or caps before each use.

Only use aerosol-tight rotor lids or caps if the seals are undamaged and clean.

Thinly brush the threads of the rotor lid screw with pivot grease (order no. Int.
5810

350.050, North America 022634330). Do not apply the pivot grease to the seals.

Replace aerosol-tight rotor lids and caps after 50autoclaving cycles.

For QuickLock rotor lids, the seal must be replaced after 50 autoclaving cycles.

Never store aerosol-tight rotors or buckets closed.

Display

Meaning

Location

Follow the instructions in the operating manual.

5418: Rear of the device
5418 R: Right side of the
device

CAUTION
Always tighten the rotor using the supplied rotor
key.

Top of device, below the
centrifuge lid

CAUTION
Close all tubes and use a rotor lid.

Top of device, below the
centrifuge lid