4 application parameters, Application parameters – Eppendorf 5418 R Centrifuge User Manual
Page 39
Technical data
Centrifuge 5418 / 5418 R
English (EN)
39
9.4
Application parameters
These values were calculated at 23°C.
Start and stop times according to DIN 58 970
Rotor
5418
5418 R
Voltage
Acceleration
times
Deceleration
times
Voltage
Acceleration
times
Deceleration
times
FA-45-18-11
230 V
120
V
100
V
16 s
16 s
18
s
18
s
18
s
18
s
230 V
120
V
100
V
13
s
13
s
13
s
13
s
13
s
13
s
run time
(5418)
30
s to 9:59
h, as well as infinity (oo),
Adjustable to 10
min in 30
s increments, then
increments of 1
min
run time
(5418 R)
0.5
min to 99
min, as well as infinity (oo),
Adjustable to 10
min in 0.5
min increments, then
increments of 1
min
Temperature
(only 5418 R)
0°C to 40°C
Relative centrifugal force (RCF)
1 to 16,873
x
g
adjustable to 5000
×
g in 100
x
g increments, then
increments of 200
×
g.
RPM
100 to 14,000
rpm
adjustable in 100
rpm
increments
Maximum load
18 micro test tubes of 2.0
mL
each.
Max. kinetic energy
2,600
Nm
Test log mandatory:
No
Allowable density of the centrifuge material (at
maximum g-force
(rcf)/speed
(rpm) and maximum
load)
1.2
g/mL