1 rcf display and calculation, Rcf display and calculation – Eppendorf 5418 R Centrifuge User Manual

Page 10

background image

Product description
Centrifuge 5418 / 5418 R
English (EN)

10

2.4.1

rcf display and calculation

To determine the g-force (rcf) for a special adapter, you can perform a calculation according to DIN 58 970
with the following formula:

rcf = 1.118 · 10

-5

· n

2

· r

max

n: Revolutions per minute (rpm)

r

max

: Max. centrifugation radius in cm

Example:

The 0.2 ml adapter has a maximum radius of 5.6 cm. At 5,000 rpm a maximum g-force of 1,565 x g is
reached.

Use the

rpm/rcf key to switch the display of centrifugation speed between rcf and rpm.

Only 5418 R: For speeds ≤ 800 rpm only the lowest adjustable g-force (100 x g) is
displayed upon switching. The exact g-force (rcf) can be determined using the formula
given below.
Ensure that the g-force (rcf) displayed upon switching is standardized to suit the rotor
without an adapter. You can achieve the following maximum g-forces (rcf) at maximum
speed (rpm) when adapters are used:

Adapter

Max. centrifugation
radius r

max

[cm]

Max. g-force (rcf)

Without adapter

7.7

16,873

for 0.2 mL PCR tubes

5.6

12,271

for 0.4 mL micro test tubes

7.7

16,873

for 0.5 mL micro test tubes

6.6

14,462

for 0.6 mL Microtainers

7.7

16,873