20

The rotor and the rotor lid must always be securely fastened.

If the centrifuge makes unusual noises when started, the rotor or rotor lid is not fastened
correctly. Switch the device off immediately by pressing "STOP".

The stainless steel rotor lid may only be used in the MiniSpin

®

/ MiniSpin

®

plus and

MiniSpin

®

SPACE centrifuges for the recommended application areas.

The rotors must only be loaded symmetrically.

Do not use centrifuges that have not been correctly installed or repaired.

Repairs must only be performed by an Eppendorf authorized service technician. Only use
original rotors and spare parts recommended by Eppendorf.

MiniSpin / MiniSpin plus may be used for the specified applications only. They must not be
operated in a hazardous or flammable environment and must not be used to centrifuge
explosive or highly reactive substances.

When handling toxic or radioactive liquids or pathogenic bacteria from Risk Group II (see World
Health Organisation: "Laboratory Biosafety Manual"), observe the safety regulations of the
country in question.

If such liquids are spilled in the rotor or rotor chamber, the centrifuge must be cleaned carefully
and properly.

Before using cleaning or decontamination methods other than those stipulated by the
manufacturer, contact the manufacturer to ensure that the intended method will not damage the
centrifuge.

Prior to centrifugation, the tubes should in any case be visually inspected for material damage.
Damaged tubes may not be centrifuged. This is because broken tubes can, in addition to
sample loss, result in further damage to the centrifuge and accessories.

Close the test tube lids before centrifuging. Open lids can be ripped off during centrifuging and
damage the centrifuge.

A liquid density of 1.2 g/ml must not be exceeded at the maximum rotational speed.

When moving the centrifuge from a cold room to a normal lab, run the centrifuge for 30 minutes
beforehand in the cold room so that it does not get covered in condensation. Alternatively, allow
it to warm up in a lab for at least three hours, but do not plug in the centrifuge in order to
prevent damage caused by condensation.

Rotors and rotor lids are high-grade components which are subject to extreme mechanical
strain. Even slight scratches and tears can lead to serious internal material damage. Avoid
damage caused by aggressive chemicals, including among others: Strong and weak alkali,
strong acids, solutions with mercury, copper and other heavy metals, chlorinated hydrocarbons,
concentrated saline solutions and phenol.

In the event of contamination caused by aggressive agents, the rotor must be cleaned immedia-
tely using a neutral cleaning liquid. This is particularly important for the bores for the tubes.

2 Safety precautions and applicational limitations

Safety precautions and applicational limitations

2