Maintenance and cleaning, Technical data – IKA HB 10 User Manual
Page 16

16
The device is maintenance-free.
Cleaning
For cleaning disconnect the main
plug.
Only use cleansing agents which have been recommended by
IKA
®
.
To clean the drive use only water with a detergent that contains
tensides, or use isopropylalcohol for stubborn soiling.
Spare parts order
When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item and designation of the spare part,
see www.ika.com, spare parts diagram and spare parts list.
Maintenance and cleaning
Repair
In case of repair the device has to be cleaned and free from any
materials which may constitute a health hazard. For this use the
"certifi cate of compliance" included or download the form from
the IKA
®
website www.ika.com.
If you require servicing, return the appliance in its original
packaging. Storage packaging is not suffi cient. Please also use
suitable transport packaging.
Technical Data
Operating voltage range
Vac
220 - 240 + 10%
100 - 120 + 10%
Rated voltage
Vac
115, 230 - 50/60 Hz
Frequence
Hz
50 / 60
Power input at nominal voltage
W
1350
Fuses
220-240 V T 10 A (5x20)
100-120 V T 16 A (5x20)
Function Heating
Heating output at nominal voltage
W
1350
Temperature range medium
°C
ambient temperature - 180 °C
Temperature indication
LCD-Display
Heating temperature setting
+1
Setting accuracy heating temperature
K
+1
Controller deviation (3l water/ 90 °C)
K
+1
Controller deviation (3l silicone oil/ 50 mPas/
K
+2
150 °C)
Adjustable safety circuit
°C
50-190
Min. volume bath container
l
1
Max. volume bath container
l
3
Materials in contact with medium
stainless steel 1.4404
Safety circuit
°C
adjustable
Dimension (W x D x H)
mm
295 x 265 x190
Weight
kg
2,8
Safety class
DIN 12877, Class 2
Protection acc. to DIN EN 60529
IP 21
Perm. ambient temperature
°C
5 - 40
Perm. humidity
%
80
Protection class
I
Overvoltage category
II
Contamination level
2
Operation at a terrestrial altitude
m
max. 2000 about sea level
Subject to technical changes!