Error message
Any malfunctions during operation will be identifi ed by an error message on the display (D), see fi g. 5.
Proceed as follows in such cases:
Switch off device using the main switch (B)
Carry out corrective measures
Restart device
Error message Cause
Effect
Correction
E 2
- In remote operation (PC), no communication
Heating off
- Switch off the device
between RV 10 and HB 10
- Clean the IR interface or
- IR connection to RV 10 disconnected
remove objects from around the IR
interface
- Switch on the device
E 3
- Temperature inside the device is too high
Heating off
- Switch off the device
- Let the device cool down
- Switch on the device
E 9
- Fault in storing the safety circuit set
Heating off
- Switch off the device
temperatures
- Let the device cool down
- Memory chip (EPROM) defective
- Switch on the device
E 21
- Safety relay does not open
Heating off
- Switch off the device and switch on again
E 23
- Adjustable safety circuit defective
Heating off
- Switch off the device
- Let the device cool down
- Switch on the device
E 24
- Set safety temperature exceeded
Heating off
- Switch off the device and switch on again
- Let the device cool down
- Switch on the device
- Check the safety temperature
setting
E 25
- The heater control circuit switch (TRIAC) is
Heating off
- Switch off the device and switch on again
faulty. The heater or the supply line is
disconnected.
E 26
- Dry running
Heating off
- Let the device cool down
- Fill up the tempering medium
- Switch off the device and switch on again
E 27
- Error in calibration
Heating off
- Switch off the device and switch on again
E 28
- Breakage of the controller sensor
Heating off
- Switch off the device and switch on again
E 29
- Short-circuit in the safety sensor
Heating off
- Switch off the device and switch on again
E 30
- Short-circuit in the controller sensor
Heating off
- Switch off the device and switch on again
E 31
- Breakage of the safety sensor
Heating off
- Switch off the device and switch on again
E 32
- Temperature deviation is too big
Heating off
- Switch off the device and switch on again
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA
®
service department,
- send the device for repair, including a short description of the fault.
HB 10.1 Spray guard
Accessories
HB 10.2 Cover (incl. spray guard)