beautypg.com

IKA KS 3000 ic control User Manual

Page 21

background image

21

Connections between shaking instrument and external

devices
PC 1.1 Cable
Cable PC 1.1 for connection to the 9-pin socket on the computer

“Watchdog” function, monitoring the serial data flow
The following applies to situations where the watchdog function
is enabled (see Namur instructions). If no new transmissions
of these commands from the PC take place within the preset
watchdog time, the heating and shaking functions will be
shutdown according to the watchdog mode selected or will be
controlled using the preset setpoints. An operating system
crash, a PC power failure or a fault in the connecting cable to
the instrument can cause an interruption in data transmission.

“Watchdog”– Mode 1
If an interruption in data transmission occurs which is longer than
the preset watchdog time, the heating and shaking functions
will be shutdown and the error message PC 1 will be displayed.

“Watchdog”– Mode 2
If an interruption in data transmission occurs which is longer than
the preset watchdog time, the speed setpoint value will be set to
the WD safety speed setpoint and the temperature setpoint will
be set to the WD safety temperature setpoint.
The PC 2 warning message will be displayed.

NAMUR Instructions Function
IN_NAME

Input description name

IN_PV_X

Reading the real value

X=1;2;3;4;
IN_SOFTWARE

Input software ID number

date, version

IN_SP_X

Reading the set rated value

X=1;2;3;4;6;12;
42;50;52;53;
IN_TYPE

Input laboratory unit ID

OUT_NAME name

Output description name. (Max. 10

characters, default: KS3000 ic)

OUT_SP_12@n

Setting the WD safety temperature with

the echo of the set value

OUT_SP_42@n

Setzen der WD safety speed with

the echo of the set value

OUT_SP_X n

Setting the rated value to n

X=1;2;4;50;52
OUT_WD1@m

Watchdog mode 1: When a WD1 event oc-

curs, the heating and shaking functions are

shutdown and message PC 1 is displayed.

Set the watchdog time to m (20...1500) se-

conds, with echo of the watchdog time.

This instruction starts the watchdog func-

tion and must be sent within the set

watchdog time

OUT_WD2@m

Watchdog mode 2: When a WD2 event oc-

curs, the speed setpoint will be set to the

WD safety setpoint speed and the tempera-

ture setpoint will be set to the WD safety

setpoint temperature. The PC 2 warning is

displayed.

The WD2 event can be reset with OUT_

WD2@0-resetting also blocks the watch

dog function. Set the watchdog time to m

(20...1500) secondes, with echo of the

watchdog time. This command starts the

watchdog function and must be sent

within the set watchdog time.

RESET

Switching off the instrument function.

START_X; X=1;2;4

Starting the instrument’s (remote) function;

(display additionally: PC)

STATUS

Display of status

1S: mode of operation A

2S: mode of operation B

3S: mode of operation C

S0: manual operation without fault

S1: Automatic operation Start

(without fault)

S2: Automatic operation Stop

(without fault)

Overview of the NAMUR-instructions
Abbreviations:
X,y =

numbering parameter (integer number)

m =

variable value, integer

n =

value of variable, floating point number

X = 1 Pt1000 medium temperature (external temperature sensor)
X = 2 temperature (incubations room)
X = 3 safety temperature
X = 4 speed
X = 6 safety speed
X = 50 Pt1000.80 medium temperature sensor

offset in K (-5.0 <= n <=+5.0)

X = 52 incubations room temperature sensor

offset in K (-5.0 <= n <=+5.0)

<0: error code: (-1)

- 1: error 1

- ... (see table)

-31: error 31

-83: wrong parity

-84: unknown instruction

-85: wrong instruction sequence

-86: invalid rated value

-87: not sufficient storage space

STOP_X

Switching off the instrument - (remote)

X=1;2;4

funktion. Variables set with OUT_SP_X are

maintained.

Contains the instruction RMP_STOP.

(display additionally: PC)

This manual is related to the following products: