Flue Gas Sample Conditioning System

Maintenance

Probe

The probe’s sponge-type filter (Figure 3, Item 9) needs to be inspected at

regular intervals for dirt contamination. Replace filter as necessary by

first pulling out the Peltier cooler from the probe and then exchanging

the filter with a new one.

Peltier Cooler

The cooler’s wadding filter (Figure 3, Item 12) needs to be inspected at

regular intervals for dirt contamination. Replace filter as necessary by

first pulling off the filter chamber’s top cap and then exchanging the

filter with a new one.

Condensate Pump

WARNING!

The condensed water will be warm and slightly

acidic. Take all necessary precautions to prevent the water

from splashing into your eyes or contacting your skin.

Empty the pump’s reservoir (Figure 3, Item 15) by first holding the

pump with its reservoir facing downwards. Then twist the reservoir 90

degrees (either clockwise or counterclockwise) to free it from the pump

base. Empty the reservoir; then reattach it to the pump. Before reat-

taching the reservoir, however, make sure that the reservoir’s o-ring

(Figure 3, Item 20) is properly positioned inside its grove in the pump

base.
The pump’s wadding filter (Figure 3, Item 17) needs to be inspected at

regular intervals for dirt contamination. Replace filter as necessary by

first unscrewing the filter chamber cap and then exchanging the filter

with a new one.