Anritsu S251B User Manual
Page 61

Step 51. Enter 9, 6, 0 using the key pad or the Up/Down arrow key. Press
ENTER
to set
the Stop Frequency to 960 MHz.
Step 52. Use the Up/Down arrow key to scroll to the
STOP FREQ LIMIT
parameter and
press
ENTER
to edit the value.
Step 53. Enter ., 5, 5. Press
ENTER
to set the Stop Frequency Limit value.
Step 54. Press
ENTER
at OK to accept the values and activate the Segment 1 limit line.
Step 55. Press the
SELECT SEGMENT
soft key and use the Up/Down arrow key to scroll
to segment #2. Press the
EDIT SEGMENT
soft key to change the parameter val-
ues for segment #2.
Step 56. Repeat the above steps as necessary to set the Segment #2 Start Frequency to
1710 MHz, Start Limit to 0.70, Stop Frequency to 1990 MHz, and Stop Limit to
0.8. Figure 4-12 depicts a typical cable loss measurement.
NOTE:
Refer to page 4-4 for instructions on saving and recalling a setup or display.
4-15
Chapter 4 Measurements
Figure 4-12.
Cable Loss Measurement Results