beautypg.com

MyWeigh 1SCALE User Manual

Page 6

background image

9. Gently Place a 200g standard weight on the platform. The displayed figure will increase.

10. Wait 10 seconds for the display to stabilize and then gently press the Big Main Button once. The

display should change to 0000.

11. Adjust the settings by using the Tare key (shifts the blinking digit) and the G/OZ key (changes the

blinking digit) to change the readings as described below:

12. Adjust the 0000 figure to "2000" by pressing the G/OZ key Twice. If you make a mistake and

press g/oz too many times you can just keep pressing it until you get it back to 2000.

13. Press the BIG MAIN BUTTON key. The display will show a 4 digit figure, probably 0983.

14. use the Tare & g/oz keys to change this figure to 1009.

A. Press the g/oz key to change the first figure to 1.

B. Then, press the Tare key once to shift the blinking digit.

C. Press g/oz to change the second figure to 0.

D. Then, press the Tare key once to shift the blinking digit again

E. Press g/oz twice to change the third figure to 0.

F. Then, press the Tare key once to shift the blinking digit

G. Press the g/oz key six times to change the fourth figure to 9. 1009 will show.

15. Press the Big Main Button again. The display will show the default setting 2 g. Do not change

this figure.

16. Press the Big Main Button again. The display will show 5. Press the Tare key several times to

change this to 1.

17. Press the Big Main Button again. The display will show the default setting 0.0. Do not change

this figure.

18. Press the Big Main Button again. The display will show 1000. Press the Tare key several times to

change this to 2000.

19. Press the Big Main Button once more. The display will show 0 and the scale will turn off in 2

seconds, which means the calibration and Advanced programing is completed.

20. Remove the 200g weight from the platform.